TRL for diagnostics: from assay design to CDSCO approval
TRL for diagnostics in India: what each level means for an IVD kit, from assay design to test licence, performance evaluation and CDSCO manufacturing licence.
Part of the guide TRL levels (1 to 9) explained: official definitions from NASA, the EU, the DoD, NTRAF and BIRAC
TRL for diagnostics tracks an in vitro diagnostic from assay to sale. An annexure on the ANRF website puts design freeze and the Form MD-12 application at TRL 4, test batches under licence at TRL 5, pilot clinical performance evaluation with MD-25 permission at TRL 6, the main evaluation at TRL 7, and the manufacturing licence at TRL 8.
Diagnostics sit awkwardly on the standard TRL scale. The scale was written for hardware systems that are built, tested and flown. A diagnostic kit is chemistry and biology in a box: an assay, reagents, a format such as a strip or a cartridge, and a claim about what it detects in which sample. Its readiness depends as much on clinical samples and regulatory permissions as on engineering.
India now has two useful references that translate TRL for diagnostics: BIRAC's TRL document for medical devices and diagnostics, and a TRL annexure for medical devices and in vitro diagnostics (IVDs) published on the ANRF website. This post maps both onto the real steps an Indian IVD goes through, including the CDSCO forms.
What does each TRL mean for a diagnostic?
The table combines the general NTRAF wording, BIRAC's device and diagnostics definitions, and the ANRF annexure's IVD milestones for a new IVD without a predicate.
| TRL | NTRAF wording | What it means for an IVD |
|---|---|---|
| 1 | Basic principles observed and reported | A biomarker or detection principle is described |
| 2 | Technology concept formulated | Assay concept and intended use drafted; BIRAC adds market and patent landscaping and FTO |
| 3 | Experimental proof of concept | The assay detects the target in the lab with spiked or limited samples |
| 4 | Component and/or system validation in lab | Design freeze, risk classification, test licence application on MD-12 |
| 5 | Laboratory scale, similar system validation in relevant environment | Test batches made under an MD-13 licence; analytical performance and stability tested |
| 6 | Engineering/pilot-scale prototype demonstrated in relevant environment | Permission for clinical performance evaluation (MD-25); pilot data on clinical samples |
| 7 | Full scale system prototype demonstration in operational environment | Main (confirmatory) clinical performance evaluation; data submitted for manufacturing permission |
| 8 | System complete and qualified through test and demonstration | Manufacturing licence obtained; production under ISO 13485 |
| 9 | Actual system proven in operational environment over full range of expected conditions | Launched, with post-market surveillance |
Sources: NTRAF, BIRAC TRL document for medical devices and diagnostics and the TRL annexure on the ANRF website. The annexure has no date or issuing body printed on it, so check the version attached to the call you are applying to.
NTRAF uses the same wording for every technology, which is why the last column translates it for a diagnostic. ZTRAF, used in TRL Pro, has separate sub-domains for Class A/B and Class C/D devices and diagnostics, so a low-risk kit and a high-risk test get different questions.
For the broader device picture, see TRL for medical devices.
TRL for a new in vitro diagnostic
Where do the CDSCO forms fall?
Under the Medical Devices Rules, 2017, IVDs are classed by risk from Class A to Class D. CDSCO's IVD page lists the forms:
| Step | Application | Licence or permission | Typical TRL |
|---|---|---|---|
| Test licence to manufacture kits for testing, evaluation or clinical investigation | MD-12 | MD-13 | Applied at 4, held at 5 |
| Test licence to import, for the same purposes | MD-16 | MD-17 | 4 to 5 |
| Clinical performance evaluation of a new IVD | MD-24 | MD-25 | Applied at 5, held at 6 |
| Manufacturing permission for a new IVD, per the ANRF annexure | MD-28 | MD-29 | Applied at 7, held at 8 |
| Manufacturing licence, Class C or D | MD-7 or MD-8 | MD-9 or MD-10 | 8 |
| Manufacturing licence, Class A or B | State Licensing Authority | State licence | 8 |
Source: CDSCO, In Vitro Diagnostics. The same page carries CDSCO's updated list of laboratories that conduct performance evaluation of IVDs, most recently updated on 29 June 2026 at the time of writing, and the ICMR-CDSCO standard performance evaluation protocols.
The practical lesson is that the test licence comes earlier than many academic teams expect. You need it to make the batches that generate your TRL 5 data, so the application belongs in the TRL 4 plan, not after it. Our post on prototyping a diagnostic kit covers the lab validation steps in detail.
What evidence does each level need?
| TRL | Evidence a reviewer will look for |
|---|---|
| 3 | Lab data showing detection of the target, with controls |
| 4 | Frozen design, intended use statement, risk class, MD-12 filed |
| 5 | Analytical sensitivity, specificity, precision and stability on test batches made under licence |
| 6 | MD-25 permission, ethics approval, pilot clinical sensitivity and specificity |
| 7 | Main study report against a reference method on the planned sample size |
| 8 | Manufacturing licence, quality system, consistent production lots |
The step from TRL 5 to TRL 6 is where many diagnostic projects stall. Analytical data on spiked samples can look excellent, and then clinical samples bring interfering substances, low-titre positives and sample handling problems nobody planned for. Budget time and samples for that step.
New IVD or IVD with a predicate
Decide early whether a predicate exists; it changes the whole plan.
How does ZTRAF check TRL for a diagnostic?
ZTRAF, the Zernyx Technology Readiness Assessment Framework, is Zernyx's own copyrighted framework, aligned with global standards and used in TRL Pro. It covers more than 60 technology domains and subdomains. TRL Pro asks structured questions for each level, records the evidence behind each answer, and returns the assessed level with the gaps to close and the next steps, with a Zernyx TRL certificate (Zernyx's own, not a government one).
If a funder asks for your level in BIRAC or NTRAF wording, state it in that wording as well.
A worked example
Illustration with invented figures. A Bengaluru startup has a lateral flow test for a tropical infection. Lab data on spiked samples is strong, and the design is frozen.
| Step | Time | Cost (₹) | TRL reached |
|---|---|---|---|
| MD-12 application and test licence | 2 to 3 months | 50,000 in documentation | 4 to 5 |
| Three test batches with stability | 4 months | 7,00,000 | 5 |
| MD-24 application, ethics approval, pilot evaluation on 200 samples | 5 months | 9,00,000 | 6 |
| Main evaluation on the planned sample size | 6 months | 18,00,000 | 7 |
| Quality system and manufacturing licence | 6 months | 12,00,000 | 8 |
| Total | About 2 years | 46,50,000 |
The startup told an investor it was at TRL 5 before the test licence was granted. Under the annexure's milestones it was at TRL 4: the design was frozen and classified, but no batches had been made under licence. Stating TRL 4 with a dated plan to TRL 5 would have been more credible.
Common mistakes
- Making evaluation batches before the test licence is in place.
- Claiming TRL 6 on analytical data alone.
- Not deciding the risk class and intended use before design freeze.
- Assuming a predicate exists without checking.
- Using spiked samples where the reviewer expects clinical ones.
My view: for diagnostics, the regulatory forms are the clearest TRL markers you have, and a team that plans its MD-12 and MD-24 applications into its TRL roadmap will make fewer claims it later has to withdraw. The hedge is that the ANRF annexure is undated and calls change, and some reviewers still use BIRAC's or NTRAF's general wording, so check which one your funder uses.
Before you state a level in a proposal, TRL Pro assesses your technology's readiness and returns a level from TRL 1 to 9 with the gaps to close. To see which diagnostic kits institutes are licensing, read diagnostic kit technologies to license.
Common questions
At what TRL does a diagnostic kit need a CDSCO test licence?
When you start making kits for testing, evaluation or clinical performance evaluation. A TRL annexure published on the ANRF website places the Form MD-12 application at TRL 4, after design freeze and classification, and the MD-13 licence at TRL 5, when test batches are made and evaluated.
Is TRL different for an IVD with a predicate?
Yes. The same annexure gives a shorter path for IVDs with a predicate: performance evaluation and substantial equivalence at TRL 6 to 7, with the manufacturing licence application, then the licence and launch at TRL 8 to 9. A new IVD without a predicate goes through pilot and main (confirmatory) clinical performance evaluation.
What does BIRAC say TRL 4 means for a diagnostic device?
BIRAC's TRL document for medical devices and diagnostics labels TRL 4 'proof of concept established' and defines it as a functional prototype integrating the modules, with safety, efficacy and performance shown in a defined lab, simulated environment or animal model.
Can an analytical study alone take a kit to TRL 6?
Usually not. Analytical sensitivity, specificity, precision and stability support TRL 5. TRL 6 in the ANRF annexure needs permission for clinical performance evaluation and pilot data on clinical samples.